5502A
Operators Manual
5-20
UUT
5502A Calibrator
Controller
SERIAL 2
TO UUT
Port
RS-232
Port
Modem Cable
IEEE-488 Cable
gjh046.eps
Figure 5-6. Verify the RS-232 UUT Port with IEEE-488 Port
Complete the subsequent procedure to verify RS-232 UUT port operation with the IEEE-
488 port with the Win32 Interactive Control utility.
1. Complete the “IEEE-488 Port Setup Procedure” in this chapter to set up the
Calibrator for GPIB operation.
2. Complete “Verify the IEEE-488 Port” to prepare the Calibrator IEEE-488 port to be
verified. Before the last step, come to this procedure and continue to Step 3 below.
3. Go to Start then to the Programs menu.
4. Select "NI-488.2M software for... (your operating system)".
5. From the NI488.2M software menu, select "Win32 interactive control".
6. A DOS window opens with a prompt: