Remote Commands
Commands
6
6-17
FORMAT
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Format command) Use with extreme care. Restores the contents of the nonvolatile
memory device to factory defaults. The memory holds calibration constants and setup
parameters. You lose all calibration data permanently. The CALIBRATION switch on
the rear panel of the Calibrator must be set in the ENABLE position or an execution error
occurs, except for FORMAT SETUP.
Parameter: ALL (replaces the whole contents with factory defaults)
CAL (replaces all cal constants with factory defaults)
SETUP (replaces setup parameters with factory defaults)
Example: FORMAT SETUP
Replace the setup parameters with the default setup values (below). (The FORMAT ALL
command is the same as FORMAT CAL and then FORMAT SETUP.) The FORMAT
SETUP command also clears the *PUD string (see the *PUD command) and SRQSTR is
set to “SRQ: %02x %02x %04x %04x” (see the SRQSTR command) and SPLSTR is set
to “SPL: %02x %02x %04x %04x” (see the SPLSTR command).
Features
Temperature
Standard
ITS-90 Display Contrast* level 7,7
Host Connection GPIB (IEEE-488) Display Brightness* level 1,0
GPIB Port Address 4 RTD Power Up
Default Type
pt385
Serial Ports 8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud
Thermocouple
Power Up Default
Type
K
EOL (end of line) CRLF Current Limits ±20.5 A
EOF (end of file) 012,000 Voltage Limits ±1020 V
Remote I/F term
Remote Commands
SRQSTR SRQ: %02x %02x %04x
%04x
*PUD string cleared
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
Defaults
dBm Impedance 600 Ω
FUNC?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Function query) Shows the current output, measurement, or calibration function. See the
response below for output and measurement modes.
Responses: DCV (dc volts function)
ACV (ac volts function)
DCI (dc current function)
ACI (ac current function)
RES (ohms function)