EasyManua.ls Logo

Fluke 5502A - Page 211

Fluke 5502A
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
Commands
6
6-19
Responses: (Indefinite ASCII) A message that has four fields separated by commas as
follows:
1. Manufacturer
2. Model number
3. Serial number
4. Firmware revision levels for the Main CPU+Front Panel CPU+Inguard
PGA
Example: *IDN? shows FLUKE,5502A,5248000,1.0+1.3+1.8
Shows Fluke manufacturer, model 5502A, serial number 5248000, main firmware
version 1.0, encoder firmware 1.3, and inguard PGA 1.8.
INCR
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Increment command) Increments or decrements the output (as selected if you use the
EDIT command, or defaults to the primary output) and records error mode; the same as if
you use the Calibrator output adjustment knob in local operation.
Parameters: <+ value> (increment value) (optional unit matching edit field)
<–value> (decrement value)
Example: INCR +.00001 mV
Load the error mode and increment the selected edit field by .00001 mV.
ISCE
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status Change Enable command) Puts two bytes into the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable Registers” in
Chapter 5 for more information.)
Parameter: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCE 6272
Load decimal 6272 (binary 0001010001000000) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT). This is equivalent to sending the commands
ISCE0 6272 and ISCE1 6272 (see below).
ISCE?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status Change Enable query) Shows the two bytes from the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable Registers” in
Chapter 5 for more information.)
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCE? shows 6272
Shows decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCE0
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status 1 to 0 Change Enable command) Puts the two bytes into the 16-bit
ISCE0 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Parameter: <value> (decimal equivalent of the 16 bits, 0 to 32767)

Table of Contents

Other manuals for Fluke 5502A

Related product manuals