Remote Commands
Commands
6
6-39
TC_TYPE_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Thermocouple Type Default query) Shows the default thermocouple (TC) sensor type.
Responses: B (B-type thermocouple)
C (C-type thermocouple)
E (E-type thermocouple)
J (J-type thermocouple)
K (K-type thermocouple) (default)
N (N-type thermocouple)
R (R-type thermocouple)
S (S-type thermocouple)
T (T-type thermocouple)
X (10 μV/°C linear output)
Y (% relative humidity)
Z (1 mV/°C linear output)
Example: TC_TYPE_D? shows K
Shows K when the thermocouple type default is a type-K thermocouple.
TEMP_STD
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Temperature Degree Standard command) Selects the temperature standard IPTS -68
(1968 International Provisional Temperature Standard) or its-90 (1990 International
Temperature Standard), which is kept in the Calibrator non-volatile memory. (While
saving configuration data in the non-volatile memory, a period of about 2 seconds, the
Calibrator does not respond to remote commands.) The default is its-90.
Parameters: IPTS_68
ITS_90
Example: TEMP_STD ITS_90
See the temperature standard to ITS-90.
TEMP_STD?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Temperature Degree Standard command) Shows the temperature standard IPTS-68
(1968 International Provisional Temperature Standard) or ITS-90 (1990 International
Temperature Standard).
Parameters: IPTS_68
ITS_90
Example: TEMP_STD? shows ITS_90
Shows ITS_90 when the temperature degree standard is the 1990 International
Temperature Standard.