Remote Commands
Commands
6
6-43
UUT_SENDB
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Send UUT Binary Data command) Send binary data to the UUT serial port (Calibrator
rear panel SERIAL 2 to UUT serial port). Use the UUT_SEND command instead of
sending ASCII data. The command can be sent over GPIB or RS-232 ports, but applies to
SERIAL 2 TO UUT serial port operation.
Parameter: Comma separated integers to send (maximum of 10)
Example: UUT_SENDB 42,73,68,78,63,10
Send the ASCII characters "*IDN?" followed by a new line (ASCII 10) to the UUT serial
port.
UUT_SET
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(UUT Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 2 TO UUT serial port, which is kept in the Calibrator non-volatile memory.
(While saving configuration data in the non-volatile memory, a period of about 2
seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 1 FROM HOST serial port, see the SP_SET command.) The
factory default values are shown below in BOLD type. (To return to the factory defaults,
see the FORMAT SETUP command.)
The interface selection sets the command response, with command echo back with
TERM (terminal) and no echo back with COMP (computer).
Parameters: <baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<flow control>, XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity> PNONE (none), PODD (odd),PEVEN (even)
Example: UUT_SET 9600,XON,DBIT8,SBIT1,PNONE
Set the parameters for the rear panel SERIAL 2 TO UUT serial port to the factory default
values.
UUT_SET?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(UUT Serial Port Set query) Shows the RS-232-C settings for the Calibrator rear panel
SERIAL 2 TO UUT serial port. (To return the parameters for the rear panel SERIAL 1
FROM HOST serial port, see the SP_SET? command.) The factory default values are
shown below in BOLD type. (To return to the factory defaults, see the FORMAT SETUP
command.)
Responses: <baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<flow control>, XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity> PNONE (none), PODD (odd),PEVEN (even)
Example: UUT_SET? shows 9600,XON,DBIT8,SBIT1,PNONE
Shows the parameters for the rear panel SERIAL 2 TO UUT serial port, as shown, when
set to the factory default values.