5502A
Operators Manual
6-46
ZERO_MEAS
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
Sets the zero offset for capacitance measurement with the -SC600.
Parameter: 1. (boolean) ON
(boolean) OFF
Example: ZERO_MEAS ON
Sets the zero offset to the current measurement value.
ZERO_MEAS?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
Shows the zero offset for capacitance measurements with the -SC600.
Parameter: (None)
Responses: 1. (character) OFF (no zero in effect)
(character) ON (zero in effect)
2. (float) offset value
3. (character) units (F)
Example: ZERO_MEAS? shows ON,-3.66E-02,F