Maintenance
Performance Tests
7
7-13
Table 7-7. Verification Tests for AC Voltage (Normal) (cont.)
Range Output Frequency Lower Limit Upper Limit
32.9999 V 30.0000 V 20 kHz 29.9802 V 30.0198 V
32.9999 V 30.0000 V 50 kHz 29.9736 V 30.0264 V
32.9999 V 30.0000 V 90 kHz 29.9404 V 30.0596 V
329.999 V 33.000 V 45 Hz 32.984 V 33.015 V
329.999 V 33.000 V 10 kHz 32.969 V 33.030V
329.999 V 300.000 V 45 Hz 299.880 V 300.120 V
329.999 V 300.000 V 1 kHz 299.880 V 300.120 V
329.999 V 300.000 V 10 kHz 299.799 V 300.201 V
329.999 V 300.000 V 18 kHz 299.754 V 300.246 V
329.999 V 300.000 V 50 kHz 299.703 V 300.297 V
329.999 V 200.000 V 100 kHz 199.536 V 200.464 V
1020.00 V 330.00 V 45 Hz 329.84 V 330.15 V
1020.00 V 330.00 V 10 kHz 329.73 V 330.26 V
1020.00 V 1000.00V 45 Hz 999.56 V 1000.44 V
1020.00 V 1000.00 V 1 kHz 999.56 V 1000.44 V
1020.00 V 1000.00 V 5 kHz 999.349 V 1000.66 V
1020.00 V 1000.00 V 8 kHz 999.23 V 1000.77 V
1020.00 V 1020.00 V 1 kHz 1019.55 V 1020.44 V
1020.00 V 1020.00 V 8 kHz 1019.21 V 1020.78 V
[1] Typical specification is -24 dB at 2 MHz
Table 7-8. Verification Tests for AC Voltage (AUX)
Range Output, AUX
[1]
Frequency Lower Limit Upper Limit
329.999 mV 10.000 mV 45 Hz 9.622 mV 10.378 mV
329.999 mV 10.000 mV 1 kHz 9.622 mV 10.378 mV
329.999 mV 10.000 mV 5 kHz 9.535 mV 10.465 mV
329.999 mV 10.000 mV 10 kHz 9.520 mV 10.480 mV
329.999 mV 10.000 mV 30 kHz 8.700 mV 11.300 mV
329.999 mV 300.000 mV 9.5 Hz 283.500 mV 316.500 mV
329.999 mV 300.000 mV 10 Hz 299.180 mV 300.820 mV
329.999 mV 300.000 mV 45 Hz 299.390 mV 300.610 mV
329.999 mV 300.000 mV 1 kHz 299.390 mV 300.610 mV
329.999 mV 300.000 mV 5 kHz 299.100 mV 300.900 mV
329.999 mV 300.000 mV 10 kHz 298.650 mV 301.350 mV