5502A
Operators Manual
7-16
Table 7-9. Verification Tests for AC Current (cont.)
Range Output Frequency Lower Limit Upper Limit
2.99999 A 0.33000 A 5 kHz 0.32735 A 0.33265 A
2.99999 A 0.33000 A 10 kHz 0.31840 A 0.34160 A
2.99999 A 1.09000 A 10 Hz 1.08827 A 1.09174 A
2.99999 A 1.09000 A 45 Hz 1.08951 A 1.09049 A
2.99999 A 1.09000 A 1 kHz 1.08951 A 1.09049 A
2.99999 A 1.09000 A 5 kHz 1.08355 A 1.09645 A
2.99999 A 1.09000 A 10 kHz 1.06320 A 1.11680A
2.99999 A 2.99000 A 10 Hz 2.98542 A 2.99459 A
2.99999 A 2.99000 A 45 Hz 2.98840 A 2.99160 A
2.99999 A 2.99000 A 1 kHz 2.98840 A 2.99160 A
2.99999 A 2.99000 A 5 kHz 2.97405 A 3.00595 A
2.99999 A 2.99000 A 10 kHz 2.92520 A 3.05480 A
20.5000 A 3.3000 A 500 Hz 3.2954 A 3.3046 A
20.5000 A 3.3000 A 1 kHz 3.2954 A 3.3046 A
20.5000 A 3.3000 A 5 kHz 3.2155 A 3.3845 A
20.5000 A 11.0000 A 45 Hz 10.9840A 11.0160 A
20.5000 A 11.0000 A 65 Hz 10.9840 A 11.0160A
20.5000 A 11.0000 A 500 Hz 10.9807 A 11.0193 A
20.5000 A 11.0000 A 1 kHz 10.9807 A 11.0193 A
20.5000 A 11.0000 A 5 kHz 10.7200 A 11.2800A
20.5000 A 20.0000 A 45 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 65 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 500 Hz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 1 kHz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 5 kHz 19.4950 A 20.5050 A
Table 7-10. Verification Tests for Capacitance
Range Output
Test Frequency
or Current
Lower Limit Upper Limit
0.3999 nF 0.2200 nF 5 kHz 0.2192 nF 0.2308 nF
0.3999 nF 0.3500 nF 1 kHz 0.3387 nF 0.3613 nF
1.0999 nF 0.4800 nF 1 kHz 0.4682 nF 0.4918 nF
1.0999 nF 0.6000 nF 1 kHz 0.5877 nF 0.6123 nF
1.0999 nF 1.0000 nF 1 kHz 0.9862 nF 1.0138 nF