5502A
Operators Manual
9-32
SCOPE WAVEGEN; OUT 1V, 1 kHz (Wave Generator, 1 V peak-
to-peak, 1 kHz.)
SCOPE VIDEO; OUT 100PCT (Video trigger; 100%
amplitude)
SCOPE PULSE; OUT 100NS,1US (Pulse output, 100ns width
pulses every 1 microsecond)
SCOPE MEASZ (Impedance measurement
mode)
SCOPE OVERLD; OUT 5V (Overload test mode; output
5V)
SCOPE?
(IEEE-488, RS-232, Sequential)
Shows the oscilloscope’s current mode of operation. Shows OFF if the oscilloscope is
off.
Parameters: None
Response: <character> (Shows OFF, VOLT, EDGE, LEVSINE, MARKER,
WAVEGEN, VIDEO, PULSE, MEASZ, or OVERLD)
TRIG
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope’s trigger output BNC.
Parameters: OFF (Turns the trigger output off.)
DIV1 (Turns the trigger output on. Frequency is the same as the
signal at SCOPE output.)
DIV10 (Turns the trigger output on. Frequency is 1/10 of the signal at
SCOPE output.)
DIV100 (Turns the trigger output on. Frequency is 1/100 of the signal
at SCOPE output.)
TRIG?
(IEEE-488, RS-232, Sequential)
Shows the output setting of the oscilloscope’s trigger.
Parameters: (None)
Response: <character> (Shows OFF, DIV1, DIV10, or DIV100.)
OUT_IMP
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope’s output impedance.
Parameters: Z50 (Programs the oscilloscope output impedance to 50 Ω.)
Z1M (Programs the oscilloscope output impedance to 1 MΩ.)
OUT_IMP?
(IEEE-488, RS-232, Sequential)
Shows the impedance setting of the oscilloscope’s output.