SC600 Oscilloscope Calibration Option
Verification Tables 9
9-37
Table 9-2. SC600 Option DC Voltage Verification (cont.)
(1 MΩ output impedance unless noted)
Nominal Value (V dc) Measured Value (V dc) Deviation (V dc) 1-Year Spec. (V dc)
-0.499 0.0002895
0.5 0.00029
-0.5 0.00029
1.35 0.000715
-1.35 0.000715
2.19 0.001135
-2.19 0.001135
2.2 0.00114
-2.2 0.00114
6.6 0.00334
-6.6 0.00334
10.99 0.005535
-10.99 0.005535
11 0.00554
-11 0.00554
70.5 0.03529
-70.5 0.03529
130 0.06504
-130 0.06504
6.599 (50 Ω) 0.0165375
Table 9-3. SC600 Option AC Voltage Amplitude Verification
(1 MΩ output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-year Spec.
(V p-p)
0.001 1000 0.000041
-0.001 1000 0.000041
0.01 1000 0.00005
-0.01 1000 0.00005
0.025 1000 0.000065
-0.025 1000 0.000065
0.11 1000 0.00015