SC600 Oscilloscope Calibration Option
Verification Tables 9
9-41
Table 9-6. SC600 Option Wave Generator Amplitude Verification (50 output impedance) (cont.)
Wave Shape
Nominal Value
(V p-p)
Frequency
(Hz)
Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
triangle 0.109 1000 0.00337
triangle 0.449 1000 0.01357
triangle 1.09 1000 0.0328
triangle 2.5 1000 0.0751
Table 9-7. SC600 Option Leveled Sine Wave Verification: Amplitude
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.005 50 kHz 0.0004
0.0075 50 kHz 0.00045
0.0099 50 kHz 0.000498
0.01 50 kHz 0.0005
0.025 50 kHz 0.0008
0.039 50 kHz 0.00108
0.04 50 kHz 0.0011
0.07 50 kHz 0.0017
0.099 50 kHz 0.00228
0.1 50 kHz 0.0023
0.25 50 kHz 0.0053
0.399 50 kHz 0.00828
0.4 50 kHz 0.0083
0.8 50 kHz 0.0163
1.2 50 kHz 0.0243
1.3 50 kHz 0.0263
3.4 50 kHz 0.0683
5.5 50 kHz 0.1103