SC600 Oscilloscope Calibration Option
Verification Tables 9
9-43
Table 9-9. SC600 Option Leveled Sine Wave Verification: Harmonics (cont.)
Harmonic
Nominal Value
(V p-p)
Frequency
Measured
Value (dB)
Deviation
(dB)
1-Year Spec.
(dB)
2nd harmonic 5.5 4 MHz -33
3rd+ harmonic 5.5 4 MHz -38
2nd harmonic 5.5 8 MHz -33
3rd+ harmonic 5.5 8 MHz -38
2nd harmonic 5.5 10 MHz -33
3rd+ harmonic 5.5 10 MHz -38
2nd harmonic 5.5 20 MHz -33
3rd+ harmonic 5.5 20 MHz -38
2nd harmonic 5.5 40 MHz -33
3rd+ harmonic 5.5 40 MHz -38
2nd harmonic 5.5 80 MHz -33
3rd+ harmonic 5.5 80 MHz -38
2nd harmonic 5.5 100 MHz -33
3rd+ harmonic 5.5 100 MHz -38
2nd harmonic 5.5 200 MHz -33
3rd+ harmonic 5.5 200 MHz -38
2nd harmonic 5.5 400 MHz -33
3rd+ harmonic 5.5 400 MHz -38
2nd harmonic 5.5 600 MHz -33
3rd+ harmonic 5.5 600 MHz -38
Table 9-10. SC600 Option Leveled Sine Wave Verification: Flatness
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.005 50 kHz na na
0.005 30 MHz 0.000175
0.005 70 MHz 0.000175
0.005 120 MHz 0.0002
0.005 290 MHz 0.0002
0.005 360 MHz 0.0003
0.005 390 MHz 0.0003
0.005 400 MHz 0.0003