SC600 Oscilloscope Calibration Option
Verification Tables 9
9-51
Table 9-10. SC600 Option Leveled Sine Wave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
5.5 580 MHz 0.2201
5.5 590 MHz 0.2201
5.5 600 MHz 0.2201
Table 9-11. SC600 Option Edge Verification: Amplitude
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.005 1 kHz 0.0003
0.005 10 kHz 0.0003
0.005 100 kHz 0.0003
0.01 100 kHz 0.0004
0.025 100 kHz 0.0007
0.05 100 kHz 0.0012
0.1 100 kHz 0.0022
0.25 100 kHz 0.0052
0.5 100 kHz 0.0102
1 100 kHz 0.0202
2.5 100 kHz 0.0502
2.5 10 kHz 0.0502
2.5 1 kHz 0.0502
Table 9-12. SC600 Option Edge Verification: Frequency
Nominal Value
(V p-p)
Frequency
Measured Value
(Hz)
Deviation (Hz) 1-Year Spec. (Hz)
2.5 1 kHz 0.0025
2.5 10 kHz 0.025
2.5 100 kHz 0.25
2.5 1 MHz 2.5
2.5 10 MHz 25