SC600 Oscilloscope Calibration Option
Verification Tables 9
9-53
Table 9-16. SC600 Option Marker Generator Verification
Period (s) Measured Value (s) Deviation (s) 1-Year Spec. (s)
5 0.0251 s
2 0.00405 s
0.05 3.75E-06 s
0.02 5E-8
0.01 2.5E-8
1e-7 2.5E-13
5e-8 1.25E-13
2e-8 5E-14
1e-8 2.5E-14
5e-9 1.25E-14
2e-9 5E-15
Table 9-17. SC600 Option Pulse Generator Verification: Period
Nominal Value
(V p-p)
Pulse Width
(s)
Period (s)
Measured
Value (s)
Deviation (s)
1-Year Spec.
(s)
2.5 8E-08 2E-06 5E-12
2.5 0.0000005 0.01 2.5E-08
2.5 0.0000005 0.02 5E-08
Table 9-18. SC600 Option Pulse Generator Verification: Pulse Width
Nominal Value
(V p-p)
Pulse Width
(s)
Period (s)
Measured
Value (s)
Deviation (s)
1-Year Spec.
typical (s)
2.5 4.0E-09 2.0E-06 6.2E-9
2.5 4.0E-09 2.0E-05 6.2E-9
2.5 4.0E-09 2.0E-04 6.2E-9
2.5 4.0E-08 2.0E-03 4.4E-8