SC300 Oscilloscope Calibration Option
Verification Tables 10
10-33
Table 10-7. Wave Generator Function Verification: 50 Ω Load
Waveform
Nominal
Value (p-p)
Frequency
Measured Value
(p-p)
Deviation
(mV)
1-Year
Spec.
Square 5.0 mV 10 kHz 0.25 mV
Square 10.9 mV 10 kHz 0.43 mV
Square 44.9 mV 10 kHz 1.45 mV
Square 109.0 mV 10 kHz 3.37 mV
Square 449.0 mV 10 kHz 13.57 mV
Square 1.1 V 10 kHz 32.50 mV
Square 2.2 V 10 kHz 66.10 mV
Sine 5.0 mV 10 kHz 0.25 mV
Sine 10.9 mV 10 kHz 0.43 mV
Sine 44.9 mV 10 kHz 1.45 mV
Sine 109.0 mV 10 kHz 3.37 mV
Sine 449.0 mV 10 kHz 13.57 mV
Sine 1.1 V 10 kHz 32.50 mV
Sine 2.2 V 10 kHz 66.10 mV
Triangle 5.0 mV 10 kHz 0.25 mV
Triangle 10.9 mV 10 kHz 0.43 mV
Triangle 44.9 mV 10 kHz 1.45 mV
Triangle 109.0 mV 10 kHz 3.37 mV
Triangle 449.0 mV 10 kHz 13.57 mV
Triangle 1.1 V 10 kHz 32.50 mV
Triangle 2.2 V 10 kHz 66.10 mV
Table 10-8. Leveled Sine Wave Function Verification: Amplitude
Nominal Value (p-p) Frequency
Measured Value
(p-p)
Deviation
(mV)
1-Year Spec.
5.0 mV 50 kHz 0.300 mV
10.0 mV 50 kHz 0.400 mV
20.0 mV 50 kHz 0.600 mV
40.0 mV 50 kHz 1.000 mV
50.0 mV 50 kHz 1.200 mV
100.0 mV 50 kHz 2.200 mV
200.0 mV 50 kHz 4.200 mV