Fault Codes A
A-3
59 File on USB device is older than installed file
60 File on USB device same as installed file (per timestamp)
61 Error unmounting USB device
62 Cannot extract ver # from line 1 of src file (config/cal)
63 Cannot extract ver # from line 1 of dest file (config/cal)
64 Kernel Datapath1 MD5 hash failed
65 Kernel Datapath2 (readback) MD5 hash failed
100-Level Faults: System
100 Low airflow
101 Cannot modify file property
102 Update execution error
103 Cannot open gpio PORT_A
200-Level Faults: 5725 Boost
200 5725 No Error
201 5725 Self-Test ROM Failure
202 5725 Self-Test RAM Failure
203 5725 Self-Test EEPROM Failure
204 5725 Self-Test Data Bus Failure
205 5725 Self-Test CLAMPS Circuit Failure
206 5725 Self-Test HVCLR Circuit Failure
207 5725 Self-Test DAC Failure
208 5725 Self-Test Watchdog Timer Failure
209 5725 Current Heatsink Too Hot
210 Output Tripped To Standby
211 5725 Current Compliance Voltage Too High
212 5725 Current Compliance Voltage Too High
213 5725 +400 V Supply Did Not Shut Off
214 5725 -400 V Supply Did Not Shut Off
215 5725 Voltage Heatsink Too Hot
216 5725 Voltage Heatsink Too Hot
217 5725 +400 V Supply Too Small
218 5725 +400 V Supply Too Large
219 5725 -400 V Supply Too Large

Table of Contents

Other manuals for Fluke 5730A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5730A and is the answer not in the manual?

Fluke 5730A Specifications

General IconGeneral
TypeMultifunction Calibrator
Resistance Range0 to 1 GΩ
AC Voltage Frequency10 Hz to 1 MHz
DisplayColor LCD
InterfacesRS-232, Ethernet
Weight18 kg
DC Voltage Range0 to ±1100 V
Dimensions172 mm H x 432 mm W x 457 mm D (6.8 in H x 17 in W x 18 in D)