INSTALLATION AND OPERATION
Command Syntax
Select Function
—
Enter Data
-
Select
Unit
Summary
COMMAND
Set Frequency
Local: CFREO]
--
DATA
—
Remote: "FR**
—
float
—
Set Amplitude
Local: CAMPL3
—
DATA
—
Remote: "AP"
—
float
—
RANGE RESOLUTION NOTES
CMHzlVJ
CkHzjmVn
CHz|uV3
“6Z''
"MZ"
"KZ"
MHZ'*
CdBCm)D
CMHziV3
CkHzImV:
CHz|uV3
"V"
"MV"
"UV"
"NV"
0.01 to 1050 MHZ 10 Hz 1,2
0.01 to 1050 MHz 10 Hz 1,3
-127
to
+13
dBm 0.1 dBm 2,A,5
0-1 uV
to
1 V
3 digits
-127
to
+13
dBm 0.1 dBm 3,4,5
0.1 uV to 1 V 3 digits
Example
Set Frequency to 10.7 MHz and
Amplitude
to
-
1.5 dBm.
Local; CFREQ]
E13 C03 C.3 111 CMHz|V3
CAMPU
Z-1 U2 C.3 C53 CdBCm)]
Remote: "Fft10.7MZ,AP-7.SDB"
Notes
1.
Frequency
ranging
occurs
at 245 and 512 MHz.
2.
FUNCTION
([FREQ]or[AMPL])remainsselecteduntilanother
FUNCTION
or [STEP], [STO], [RCL], or [SPCL] is
pressed.
3. float equals floating-point number.
4. Amplitude uncalibrated range from -147.4
to
-127,1
dBm and from +13.1 to
+19
dBm.
5.
Amplitude
ranging
occurs
at 1/2V, 1/4V, 1/8V, ...
1/
223V with AM off and
1/4V, 1/8V,
1/16V,
...
1/224
V with AM on.