INSTALLATION AND OPERATION
Command
Syntax
Select Relative Function
Enable or Disable
Summary
COMMAND NOTE
Frequency
Local: CSPCLT
Z21
CO:]
or CIU 1
Remote: "SP"
"2"
"O'' or
"1"
1
AmpLi tude
Local:
CSPCLT C33
EO: or [1] 1
Remote: "SP"
"3"
~
"0"
or
"1"
1
Example
Set the amplitude
to -15
dB^V; i.e.,
15 dB
below 1 microvolt.
Local: LAMPU
C13 CHzluV3
LSPCU
C33 C13
LAMPL:1
Z~2 E13 C53
CdBCm)]
Remote: "AP1UV,SP31,AP-15DB"
Note
1.
1
enables the relative function; 0 disables the relative function.
Related Operations
Amplitude and
Frequency Entry
Bright-Digit
Edit
Entry
Step Entry
2-29. RF OUTPUT ON/OFF Entry
The following information describes how to enable the output of the Generator using the
RF
OUTPUT [ON/OFF] key and the corresponding remote code.
Command Syntax
RF
Output
On/Off
Summary
COMMAND NOTE
RF Output On
Local: RF OUTPUT C0N/0FF3 when 'RF OFF' is on 1
Remote; "ROl"
1
RF Output Off
Local:
RF
OUTPUT EON/OFF] when 'RF OFF' is
off
Remote; "ROO"
2-23