Maintenance
PERFORMANCE TEST
Resistor Calibrator
DC Current Source
INSTRUMENT
TYPE
DC Calibrator
Oscilloscope
Digital
Mukimeter
AC Calibrator
-
-
-
Table
6-1.
Recommended Test Equipment
MINIMUM SPECIFICATIONS
-
PREFERRED:
Voltage Range: 0-1000V dc
Voltage Accuracy: 10 pprn
Absolute Linearii:
2
1 -0 pprn
RECOMMENDED MODEL
Fluke 5700A or Fluke 5440A
ALTERNATIVE:
Fluke
343A
(Must
be
used
with
Kelvin-
Varley Voltage Divider)
Voltage Range: 0-1000V dc
Voltage Accuracy: 20 pprn
+
20 ppm of
Resistance Accuracy: 0.005%
I
Fluke 5700A or Fluke 5450A, ESI DB62
Accuracv: 20.025%
I
Fluke 5700A or Fluke 51008
Kelvin-Varley Voltage Divider:
Ratio Range: 0-1.0
Absolute Linearity: 21 ppm of input
at
dial setting
General purpose, 60 MHz, with 10
MQ
probe
1
Philips 3055 or 3355
Fluke 720A
Voltage Accuracy:
0.01% in V
dc
1.0% for 1V in V ac
@
100
kHz
Input Impedance:
10
MS2
or greater in V dc;
1 MQ or greater in parallel with <I00 pF in
V
ac
-
Fluke 8840A (with Option -09)
Fluke 5700A and Fluke 5725A
I
I
Minimum Required Accuracy (By Range)
I
I
I
1.
=(%
of setting
+
CiV)
2.
+I-(%
of setting
+
%
of range)
I
Frequency
Range
2OHz-30Hz
30 Hz
-
20 kHz
20 kHz
-
50 kHz
50 kHz
-
100 kHz
I
AC Current Source Fluke 5700A or Fluke 51008
1, 10, 100 mv'
.1
+
.005
-02
+
10
-05
+
20
-05
+
20
I
I
30
HZ
-
1 kHz
I
1
1, 10, 100v2
.1
+
.005
-02
+
.002
-05
+
-005
-05
+
-005
Frequency
Range
I
I
Construction: Soldered (not rivetted)
I
1000v2
.12
+
.005
.04
+
.OM
-08
+
.005
-1
+
.01
Minimum Required Accuracy
(All Ranges)
1
kHz
-
5 kHz =(.07%
+
1
mA)
X
frequency in kHz
I
Shorting Bar
I
Resistance: 4.5 mR
I
Pomona MDP-S-0
6-Inch Jumper
0
p
t
i
o
n a
l
T
e st
Equipment
----
I
E-Z-Hook 204-6W-S or equivalent
901 OA, 9005A or Micro-System Troubleshooter; 9000A-8048 Interface Pod;
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8840A and is the answer not in the manual?

Fluke 8840A Specifications

General IconGeneral
Frequency Range20 Hz to 100 kHz
Operating Temperature0°C to 50°C
DisplayVacuum Fluorescent
DC Current Range0 to 2 A
AC Current Range0 to 2 A