Operating instruc8ions 
OPERATING 
FWTURES 
OVER 
1 
I 
OVER: 
input 
22V 
dc 
in 
2V 
range. 
I 
Figure 
2-8. 
Overrange Indication 
In  the  external  trigger  mode,  pressing  any  front  panel 
button blanks the numeric field on the display until a new 
measurement  is  triggered. This  ensures that  all  readings 
correspond  to  the  instrument  configuration  indicated by 
the display annunciators. The blanking also occurs in the 
continuous trigger mode, but  usually isn't  noticed because 
new  measurements  are triggered  automatically. 
The TRIG button does not trigger  readings in the continu- 
ous trigger mode.  However, it does blank  the last reading 
to  acknowledge  a  button  was  pressed. 
2-20.  Reading Rates and  Noise Rejection 
The RATE button allows you  to optimize either measure- 
ment speed or noise rejection. The 8840A uses both analog 
and digital filtering to allow measurements in the presence 
of  unwanted  environmental  noise  (especially  line-related 
noise).  However,  since  filtering  introduces  a  delay  in 
response  to  a  change  in  the  input  signal,  there  is  an 
inherent  trade-off  between  noise  rejection  and  measure- 
ment  speed. 
The  instrument  has  three  reading  rates:  slow 
(S) 
and 
medium 
(M), 
with a 
5-112 
digit display, and fast 
(F), 
with 
a 4-112  digit display. To provide optimum combinations of 
measurement speed and noise rejection, the 
RATE 
button 
allows  control  of  both  the  internal  trigger  rate  and  the 
degree of filtering. The same degree of  filtering is used in 
both  the  continuous  and  external  trigger  modes. 
In 
the  continuous  trigger  mode,  the  actual  number  of 
readings  displayed  per  second  for  each  reading  rate  is 
determined by  the line-power frequency. At  power-up, the 
8840A  senses  the  line-power  frequency  and  adjusts the 
analog-to-digital converter timing characteristics for  opti- 
mum  normal-mode  noise  rejection. The resulting reading 
rates  are  shown  in  the specifications in 
Section 
1. 
2-21.  Automatic Settling Time Delay 
When  the  external trigger  mode  is  selected,  the 
8840A 
automatically inserts 
a 
delay after receiving 
a 
trigger sig- 
nal, but  before starting the first input sample. The delay is 
just  long enough so that the reading will be correct (within 
a specified number of counts of the final value) even if the 
trigger  signal  occurs 
as 
the  input  makes  a  step  change 
between zero and full scale 
(10,000 
counts and full scale 
in  the  ac  functions).  For  resistance  measurements, 
the 
delay  also  allows  for  settling when  the  step  change  is 
made  from 
an 
open  circuit  to  an  in-range value. 
The length of  the  delay depends on  the range,  function, 
and 
reading rate, as shown in the specifications in  Section 
1. 
The delay is enabled only in the external trigger mode. 
It 
can 
be  turned  off  with  a  remote  command  over 
the 
IEEE-488  interface  bus  to  accommodate  special  timing 
considerations. 
2-22.  External Trigger lnput (Option -05 Only) 
The  rear  panel  EXT  TRIG  input  is a  ?TL-level  input 
which 
can 
be 
used  to  trigger  measurements  when  the 
8840A is in  the external trigger mode.  A measurement 
is 
triggered on the falling edge of  the input. Since the 
EXT 
TRIG 
input is pulled high internally,  it  can also be  con- 
trolled  by  a  normally  open  switch. 
A 
measurement  is 
triggered when  the  switch  is closed. 
For  special applications using the IEEE-488 Interface, the 
automatic setting time delay 
can 
be disabled using remote 
commands. (See Section 
3.) 
Refer to Section 
1 
for timing 
details. 
The polarity 
of 
the EXT TRIG  input 
can 
be  reversed 
by 
changing internal jumpers.  Refer  to the Maintenance 
sec- 
tion  for  instructions. 
2-23.  Sample Complete Output (Option 
-05 
Only) 
The SAMPLE COMPLETE output indicates when  analog 
input sampling for a reading is completed. The output is a 
TTL-level  signal which  is pulsed  low  for  approximately 
2.5 
p 
when the input-sampling portion of the 
AD 
conver- 
sion is completed. The signal is useful for interfacing with 
other  equipment  when  the  8840A  is  used  in  external 
trigger  mode in  an instrumentation system. For  example, 
the  SAMPLE  COMPLETE  output  could  be  used  to 
advance  a  scanner  to the  next  channel. 
2-24.  MAKING MEASUREMENTS 
2-25.  lnput Overload Protection Limits 
WARNING 
TO  AVOID  SHOCK  HAZARD 
AND/OR 
INSTRUMENT  DAMAGE,  DO  NOT 
APPLY  INPUT  POTENTIALS  THAT 
EXCEED  THE  INPUT  OVERLOAD 
LIMITS SHOWN  IN TABLE 2-2. 
The 8840A 
is 
protected against input overloads up to the 
limits  shown  in Table 
2-2. 
Exceeding  these limits  may 
damage the  instrument  andlor pose  a  shock  hazard. 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com