Maintenance 
CAUBRATION 
Press the range button for the range to be  calibrated. 
The 8840A then  prompts for a low reference source 
for  that  range.  (See  Table 
6-12.) 
(Dufing  High- 
Frequency  AC  Calibration, the  8840A prompts  only 
for  a high  reference source.  In  this case, proceed to 
step 4.) 
Apply  the  requested  reference  source  and  press 
STORE.  The  display  will  blank  briefly  and  then 
prompt for a high  reference source. (See Table 6-12.) 
Apply  the  requested  reference  source 
and  press 
STORE.  The  display  will  blank  briefly,  and  the 
8840A will then  begin taking readings in the selected 
range so that you may verify the calibration. The CAL 
annunciator remains  on. 
To  continue, select  another  range.  You  may  restart 
any  of  the  calibration  procedures  by  pressing  the 
appropriate function button, the 
A/D 
button, or the 
HF 
AC  button. 
6-18. 
VERIFYING CALIBRATION 
When  you  complete certain 
parts 
of  the calibration proce- 
dure, the 8840A automatically begins  taking readings  so 
that you can verify the calibration is correct. (It is recom- 
mended that you  do  so.) The 
CAL 
annunciator remains lit. 
The  8840A  continues  to  take  readings  until  you  select 
another  calibration procedure  or  exit the calibration mode. 
Specifically, the 8840A begins taking verification readings 
after: 
1. 
Completing 
AD 
Calibration. 
2. 
Completing Offset  and  Gain  or  High-Frequency  AC 
Calibration. 
3. 
Completing the calibration of  an  indwidually selected 
range. 
While  the  8840A  is taking  verification readings, certain 
buttons  are  active or  function  differently: 
If  you  just  completed the  Offset 
and 
Gain 
or  High-. 
Frequency  AC Calibration for an entire function (not 
just  one  range),  the  range  buttons 
can 
be  used  to 
change  ranges  in  order  to  verify  all  ranges  were 
calibrated correctly. 
If  you  just  calibrated 
an 
individually selected range, 
pressing  another range  button 
begins 
the Offset  and 
Gain  Calibration procedure  for  the  new  range. 
You  can use the 
RATE 
button to verify the calibration 
at  other  reading  rates. 
CAUTION 
It 
is 
still possible to erase the calibra- 
tion memory while the 
8840A 
is taking 
verification  readings. 
6-'I 
9. 
ERASING CALIBRATION MEMORY 
The 8840A allows you  to erase some or  all of  the correc- 
tion  constants stored in calibration memory  (U220). It  is 
recommended that you erase the entire calibration memory 
before beginning calibration if  the calibration memory 
is 
replaced or accidentally altered. The capability of  erasing 
particular 
parts 
of  the  memory  is  mainly  intended as a 
troubleshooting aid  to  the technician. 
CAUTION 
Once the calibration memory  is erased, 
the 
8840A 
must 
be  recalibrated. 
To  erase all or 
part 
of  the calibration memory, proceed 
as 
follows: 
1. 
Press  the  front  panel 
ERASE 
button.  The  display 
should show the erase prompt "cl"  (for "clear"). 
If 
you 
do not  press another button  within  1-112 seconds, the 
8840A  returns to  its previous  state. 
2. 
To  complete an  erasure,  press  one of  the following 
buttons within 
1-112 
seconds of  pressing the ERASE 
button: 
Table 
6-12. 
Prompts 
When 
Calibrating  Individual  Ranges 
I 
AC 
Calibration 
I 
I 
I  I 
PROCEDURE 
offset 
and  Gain 
Calibration 
High-Frequency 
1 
NOTES: 
I 
1 
1. 
Exception: 
The 
1000V 
dc  range 
has 
a  high  prompt  of 
1000V 
dc. 
I 
FUNCTION 
VDC 
k& 
VAC 
(Not applicable) 
2. 
Exception: 
The 
700V 
ac range 
has 
a 
low  prompt  of 
100V 
ac and  a  high  prompt  of 
500V 
ac. 
3. 
Exception: 
The 
700V 
ac range  has  prompt  of 
200V 
ac. 
LOW  PROMPT 
Zero 
Zero 
5% 
of  full  scale2 
(No  low  prompt) 
HIGH  PROMPT 
95% 
of  full  scale1 
50% 
of  full  scale 
50% 
of  full  scale2 
50% 
of  full  scale3 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com