Maintenance 
CAUBRATlON 
each function. To save time, the 8840A uses each input for 
as 
many  ranges  as  possible. 
A 
function  is  calibrated  by  pressing  the  corresponding 
function button.  Once  a  function  is selected, the 
8840A 
automatically steps through  each  range  of  that  function, 
prompting  you  for  the necessary  reference  sources. (The 
prompts are shown  in Table 
6-10.) 
The 
8840A 
does not 
automatically select another function after one function has 
been  completely  calibrated,  therefore,  the  functions may 
be  calibrated  in  any  order. 
To  perform  Offset  and  Gain  Calibration, proceed 
as 
fol- 
lows: 
1. 
Ensure the 
AD 
Calibration procedure has been  com- 
pleted. 
2. 
Select  the  desired  function  by  pressing  the  corre- 
sponding function button.  The 8840A will display the 
fust  prompt  for  that  function. 
3. 
Each  time  the 
8840A 
prompts  you  for 
a 
reference 
source, apply this source to the appropriate terminals, 
and  press  the  STORE  button.  When  STORE  is 
pressed,  the  numeric  display  field  blanks  while  the 
8840A performs the  necessary  calculations. (Do  not 
change  the  reference  source  while  the  display  is 
blank.) The 
8840A 
then  displays the next prompt. For 
reference,  all prompts are shown in  Table 
6-10. 
NOTE 
To 
use 
reference sources that  dfler from  the 
prompted  values, see  Storing  Variable Inputs 
later  in  this section. 
4.. 
After  the  last  range  is  calibrated,  the 8840A begins 
taking readings in the highest  range so that you  may 
verify  its calibration. (The 
CAL 
annunciator  remains 
on.)  To  verify  the  calibration for  the  other  ranges, 
press  the  corresponding  range  button.  (Pressing  a 
function button begins the Offset and Gain Calibration 
procedure  for  that  function.) 
5. 
Repeat steps 
2, 
3, 
and 
4 
for the remaining functions. 
Note that 
both 
2-wire 
and 
4-wire ohms functions must 
be  calibrated.  (The  VAC  and  mA  AC  functions 
require calibration only 
if 
the True 
RMS 
AC 
option is 
installed.) 
NOTE 
(True 
RMS 
AC  option only.) The  VAC 
and 
mA 
AC 
fwrctions  should  normally 
be 
calibrated 
using reference sources at 
I 
Wlz 
(210%). 
For 
special  applications,  performance  may  be 
optimized at other frequencies.  See Optimizing 
AC 
Calibration 
at 
Other Frequencies, later in 
this  section. 
6.. 
When 
all 
functions  have  been  calibrated,  exit  the 
calibration  mode  by  pressing  the  CAL 
ENABLE 
switch  and  attach  a  calibration  certification  sticker 
over the 
CAL 
ENABLE 
switch. 
(If 
the True RMS AC 
option  is  installed,  instead  proceed  to  the  High- 
Frequency  AC  Calibration procedure which  follows.) 
6-1 
4. 
HIGH-FREQUENCY 
AC 
CALIBRATION 
The High-Frequency  AC Calibration procedure calibrates 
the  response of  the  VAC  function from  20 
kHz 
to 
100 
Hz. 
If  the True RMS AC option is not installed, selecting 
this  procedure  results  in 
an 
error  message. 
Table 
6-10. 
Offset 
and Gain Calibration 
DISPLAYED 
PROMPT 
2 
WlRE 
162 
VDC 
I 
vAcl 
I 
4 
WlRE 
162 
+00.0 
mV 
DC 
(short) 
+190.0 
mV 
DC 
+I 
900 
V 
DC 
+19.00 
V 
DC 
+190.0 
V 
DC 
+1000. 
V 
DC 
0.00 R 
(short) 
100.0 
9 
1.000 kQ 
10.00 kQ 
1 
100.0 
kR 
Step 
G 
not 
applicable 
for  these functions. 
Steps 
C-G 
not  applicable 
for these functions. 
STEP 
A 
6 
C 
D 
E 
F 
G 
10.00 
MQ 
- 
- 
NOTE: 
1. 
ItIp~ts should  be at 
1 
kHz 
+/-lo%. 
Performance 
may 
be 
enhanced for  specific frequencies (see 
text). 
I 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com