Maintenance Test 
PERFORMANCE 
TEST 
Ensure the 8840A is on and has warmed up  for at least 
1 
hour. 
Select the VDC function. 
Connect the 
DC 
Calibrator (see Table 6-1) to provide 
a voltage input to the HI  and 
LO 
INPUT 
terminals. If 
using the Fluke 343A and the  Kelvin-Varley Voltage 
Divider instead of  the Fluke 5440A, connect the test 
leads as shown in Figure 6-1. 
For each step in Table 6-2, select the indicated range, 
set  the  DC  Calibrator  for  the  specified  input,  and 
verify that the displayed reading  is  within  the  limits 
shown for  each  reading rate.  (For  step A,  connect a 
short  across  the  HI  and 
LO 
INPUT 
terminals  and 
press OFFSET. The measurement in step C should be 
relative to this offset.) 
Set the DC Calibrator to  input negative voltages, and 
repeat steps C through 
G 
of Table 
6-2. 
With the unit is the 2V range, check the 
AID 
linearity 
by  setting the DC Calibrator for each step in Table 6- 
9, 
while  verifying the  display  reading  is  within  the 
limit shown. Set the DC Calibrator for zero volts and 
disconnect if from the 8840A. 
6-5. 
AC 
Voltage 
Test 
(Option 
-09 
Only) 
The  following  procedure  may  be  used  to  verify  the 
accuracy of the 
VAC 
function: 
Ensure the 8840A is on  and warmed up for at  least 
1 
hour. 
Select  the  VAC  function  and  the  slow 
(S) 
reading 
rate. 
Connect the AC  Calibrator to provide 
a 
voltage input 
to the HI and LO INPUT terminals. 
(Low-  and  Mid-Frequency  Test.)  For  each  step  in 
Table  6-3,  select  the  indicated  range,  set  the  AC 
Calibrator for the  specified input, and  verify that  the 
displayed reading is within the limits shown for each 
reading rate. 
NOTE 
This procedure  tests  the extremes of  each range. 
You may shorten the procedure  by testing only the 
"quick  test points"  indicated  in  Table 
6-3 
with 
asterisks. 
(High-Frequency Test.)  for  each  step  in  Table 
6-4, 
select the indicated range, set the AC  Calibrator for 
the  specified  input,  and  verify  that  the  displayed 
reading  is  within  the  limits shown  for  each  reading 
rate. 
NOTE 
This procedure  tests the extremes of  each range. 
You may shorten the procedure by testing only the 
"quick  test points"  indicated 
in 
Table 
6-4 
with 
asterisks. 
DC CALIBRATOR  S!NSE 
- 
(FLUKE 343A) 
f--8 
t-t- 
REVERSE THESE 
1 
I 
LEADS FOR 
NEGATIVE INPUT 
VOLTAGE TO 8840A 
DIVIDER (FLUKE 720A) 
OUTPUT 
I 
3-b~~o~ 
ooo 
ad+ 
I 
CONNECTIONS ARE SHOWN 
FOR POSITIVE INPUT  VOLTAGES 
TO 
8840A 
NOTE: THE KELVIN-VARLEY VOLTAGE DlVlDER 
IS 
NOT REQUIRED FOR INPUT VOLTAGES ABOVE 2.OV  DC 
8840A 
- 
- 
- 
Figure 
6-1. 
Connections for Kelvin-Varley Voltage Divider 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com