Maintenance 
CALIBRATION 
installing the True RMS AC option, it  may  be  necessary 
only  to  perform  Offset  and Gain  Calibration for  the  ac 
functions,  followed  by  High-Frequency 
AC 
Calibration. 
But  if  the 
A/D 
Calibration is performed,  it  must  be  fol- 
lowed by  a complete  Offset and Gain  Calibration for all 
functions and  then  by  High-Frequency 
AC 
Calibration. 
Some of  the calibration calculations are complex and take 
the 
8840A 
some time  to execute. For example, when  you 
store the zero input during the Offset and Gain Calibration 
for  the 
VDC  function,  it takes around 
22 
seconds before 
the  next  prompt  appears. (The 
8840A 
automatically  uses 
this input to calibrate the offset for all ranges.)  While the 
8840A 
is executing 
a 
calibration step, it ignores all of  the 
front  panel  buttons  and  delays  execution  of  dl remote 
commands. 
6-1 
1. 
INITIAL 
PROCEDURE 
Always  begin  the  calibration procedure  as  follows: 
Allow the 
8840A 
to stabilize in an  environment with 
ambient  temperature  of 
18°C 
to 
28OC 
and  relative 
humidity  less  than 
75%. 
Turn 
the 
8840A 
on  and allow  it  to  warm  up  for  at 
least 
1 
hour. 
Enable  the  calibration  mode  by  pressing  the 
CAL 
ENABLE  switch  with  a small  screw-driver  or  other 
suitable  instrument. 
(The 
CAL  ENABLE  switch  is 
located on the right side of  the display and is normally 
covered by  a  calibration seal.) 
When the calibration mode is enabled, the 
CAL 
annuncia- 
tor lights up,  and the 
8840A 
displays the first prompt for 
the 
A/D 
Calibration procedure (Figure 
6-2). 
To  exit  the 
calibration mode, press  the 
CAL 
ENABLE  switch again. 
In  the calibration mode,  the front  panel  controls assume 
the  functions  described  in  Figure 
6-3. 
Some  of  these 
functions are advanced features and  are  not  required for 
the basic calibration procedure. The display blanks briefly 
when  a  button 
is 
pressed. 
The following functions are inappropriate during calibra- 
tion,  and  are  therefore  unavailable: 
Offset 
- 
Autoranging 
External Trigger 
- 
-- 
-- 
Figure 
6-2. 
First 
ND 
Calibration 
Prompt 
Front  Panel  Trigger 
Front  panel 
SRQ 
(under  local  control) 
Diagnostic  self-tests 
6-12. 
A/D 
CALIBRATION 
The 
A/D 
Calibration procedure  calibrates the  analog-to- 
digital converter for offset, gain and linearity. The 
8840A 
automatically selects the 
A/D 
calibration procedure  when 
the CAL 
ENABLE 
switch is first  pressed. The procedure 
must 
be 
performed  in  its  entirety,  and may  not 
be 
per- 
formed in part. If  the 
AID 
calibration is discontinued prior 
to  completion,  the  last  complete  set  of 
AD 
calibration 
constants will  be  retained  unchanged. 
perform 
AD 
Calibration, proceed 
as 
follows: 
Ensure the Initial Procedure has been completed. The 
8840A then displays the prompt  for the first reference 
source,  zero  volts  (i-e., a  short). 
Each  time  the 
8840A prompts  you  for  a  reference 
source, apply the requested source to the HI 
and 
LO 
INPUT 
terminals, and press the STORE button. When 
STORE is pressed,  the numeric  display  field blanks 
while the 
8840A 
performs the necessary calculations. 
(Do not change the reference source while the display 
is  blank.)  The 
8840A 
then  displays the next  prompt. 
For  reference,  all  prompts  are  shown  in  Table 
6-8. 
NOTE 
The 
8840A 
automatically checks that the refer- 
ence  input 
is 
near 
the 
value  prompted, 
and 
displays 
ERROR 
41 
if 
it  exceeds 
a 
specific 
tolerance. (See Advanced Features 
and 
Special 
Considerations, later  in 
this 
section.) 
After the last input is stored, the 8840A begins taking 
readings in the 
2V 
range of  the VDC  function.  (The 
CAL annunciator remains on.) Verify the 
A/D 
calibra- 
tion using the test points in Table 
6-9. 
If  you wish to 
repeat  the 
AID 
Calibration procedure, press  the AID 
button. 
The 
A/D 
Calibration  procedure  is  an  iterative  process. 
Each pass through the procedure uses the constants stored 
previously  and  improves  them.  Normally,  one  pass  is 
adexpate.  However,  if  the  calibration  memory 
has 
been 
erased  or  replaced, or  the 
AD 
Converter has  undergone 
repair, the 
AID 
Calibration procedure must 
be 
performed 
tw~ce. 
Because  the 
A/D 
Calibration  procedure  is iterative, the 
instrument's  performance  can  be  slightly  enhanced  by 
going through the  procedure  more  than  twice.  However, 
this is not  necessary to meet  the published specifications. 
6-13. 
OFFSET 
AND 
GAIN 
CALIBRATION 
This  procedure  calibrates  the  instrument's  offsets  and 
gains by  applying 
a 
high 
and low  input for every range of 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com