~ectio) 
4 
Measurement 
Tuto 
ial 
~ 
4-1. 
INTRODUCTION 
This  section  discusses  considerations  and  techniques  to 
Inpur, protecfion circuitry can reduce the inpd 
help  you  use the 8840A effectively. Among  other things, 
impedance to 
as 
low 
as 
100 
kQ 
when the in 
this section discusses sources of  error which  are an inher- 
is 
overrange. 
This 
may 
also 
occur momentaril 
I 
ent 
part 
of  the measurement process and which  occur for 
when  the  instrumenf auoranges  to  a  highe 
all  multimeters. By  understanding  why  and  when  these 
range. 
t 
errors occur, and by  knowing how  and when to correct for 
I 
them,  you 
can 
make  accurate  measurements  with  confi- 
dence. 
44. 
Inpqt 
Bias 
Current 
Error 
I 
I 
Input bias current error occurs because a voltmeter7 
This section also discusses the relative benefits of 2-wire 
bias 
curreht  always  changes  the  voltage  of  the 
and 4-wire ohms, describes special considerations for mak- 
under  test.  However,  the  error  is  significant  only when 
ing  ac  measurements,  and  presents  some  unusual 
measuring  voltages  in  circuits  with  very  high 
applications--for  example,  using  the  test  current  in  the 
inpdance. The error 
can 
be 
measured 
as 
shown in 
2-wire ohms function 
as 
a 
troubleshooting tool  in  itself. 
4-2. 
4-2. 
DC 
VOLTAGE MEASUREMENT 
With the &840A, it  is easy  to correct for this errod using 
When  measuring 
dc 
voltages in  high-impedance circuits, 
the  OFFSET button: 
I 
there are two possible sources of  error to consider: circuit 
1.  Select  the 
VDC 
function and  the  desired 
loading  and  input  bias  current. 
2. 
Come# 
the  8840A 
INPUT 
terminals  to 
4-3. 
Circuit 
Loading 
Error 
which matches the source impedance of  the 
Whenever a voitmder is connected to a circuit, the volt- 
be 
tested. 
meter's  internal  resistance  changes  the  voltage  of  the 
3. 
MOW 
the  displayed  reading  senle. 
circuit  under  test.  The  resulting  error  is  called  circuit 
I 
loading error. The error is negligible 
as 
long 
as 
the resis.- 
4- 
Press 
the 
OFFSET 
button. 
I 
tance  of  the circuit under  test  (the  source  impedance) is 
5. 
Remove  the  resistor. 
small compared to the  input  impedance of  the meter. 
As 
I 
the source impedance approaches the input  impedance of 
6. 
Proceed 
with the  desired  nmsu~ement- 
the voltmeter, the error can  be  considerable. The percent- 
, 
age of  error can 
be 
calculated using the formula in  Figure 
4-1. 
Measure  a  1.5V  source  with 
1 
MQ 
source impcidance, 
The input impedance of the 8840A is 10 
MQ 
in  the 200V 
correcting 
for 
input 
bias 
current. 
and  lOOOV  ranges, and is  greater than  10,000 
MQ 
ia the 
connect 
a 
M~ 
rsistor 
between 
the 
l~PU~ 
bI 
md 
200  mV,  2V,  and 20V ranges.  Therefore, for the 8840A, 
circuit  loading  error  is less than  0.01%  as  long  as  the 
INPUT 
LO 
terminals. 
I 
source impedance is less than 
1 
MQ 
in  the 200 mV,  2V7 
2. 
Select  the  VDC  function and  the  2V  range. 
and  2OV  ranges,  and  less  than 
1 
kQ 
in  the  2OOV  and 
Allow 
the 
display 
to 
setile. 
I 
l0OOV  ranges. The exceptionally high input impedance on 
I 
the 20V dc range allows high-accuracy readings in CMOS 
4. 
Press  OFFSET.  (This  zeroes  the  input  bias  current 
and  high-impedance  analog  circuitry. 
error 
.) 
I 
I 
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com