Maintenance 
CALIBRATION 
Table 
6-15. 
Error 
Numbers 
Which 
Are  Displayed 
When 
Commands  Are  Not  Valid 
COMMAND  NORMAL 
MODE 
SELF 
TEST 
I 
CALIBRATION MODE 
AID  CAL  OFFSET 
& 
GAIN CAL 
HF AC 
CAL 
Although not usually necessary, these methods 
can 
be used 
for  other  commands 
as 
well. 
6-25. 
REMOTE 
ERASURE 
The C3 command allows you to erase the entire calibration 
memory.  The  erasure  is  executed  by  sending the string 
"C3  CO"  (equivalent  to  pressing  ERASE  and  then 
STORE). 
Any 
command other than 
CO 
after C3 will abort 
the erasure. To facilitate remote  calibration, the 
CO 
com- 
VERIFICATION 
=I 
mand  does not  timeout 
as 
does the  front panel 
ERASE 
button. The selective erasure that is possible from the front 
panel  is  intended 
as 
a  troubleshooting  aid,  and  is  not 
available over  the  IEEE-488 Interface. 
Note that the erase command can take up to 3 seconds to 
execute. To prevent timeout problems with the controller, 
you  must  determine  when  the  command  is  completed 
before continuing. Several methods are presented in Tirn- 
ing  Considerations,  above. 