Calibration Adjustment
Final Calibration For v11.10 and later 5
5-17
Table 5-3. Input C LF-HF Gain Calibration Points
Cal step UUT input signal 5502A Setting
CL 0694 0.5 Vpp square wave, 1 kHz SCOPE edge, 0.5 Vpp, 1 kHz
CL 0455 100 mVpp square wave, 1 kHz SCOPE edge, 100 mVpp, 1 kHz
CL 0550 2.5 Vpp square wave, 1 kHz SCOPE edge, 2.5 Vpp, 1 kHz
CL 0584 2.5 Vpp sine wave, 50 kHz SCOPE levsine, 2.5 Vpp, 50 kHz
CL 0585 2.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504: 501 MHz
Fluke 190-202/204: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 2.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0484 500 mVpp sine wave, 50 kHz SCOPE levsine, 500 mVpp, 50 kHz
CL 0485 0.5 Vpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504: 501 MHz
Fluke 190-202/204: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
Fluke 190-202/204: 221 MHz
Fluke 190-102/104: 141 MHz
Fluke 190-062: 91 MHz
SCOPE levsine, 0.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
221 MHz
141 MHz
91 MHz
CL 0464 100 mVpp sine wave, 50 kHz SCOPE levsine, 100 mVpp, 50 kHz
CL 0465 100 mVpp sine wave
Red subversion 5; Blue Subversion 2
Fluke 190-502/504: 501 MHz
Fluke 190-202/204: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
Fluke 190-202/204: 221 MHz
Fluke 190-102/104: 141 MHz
Fluke 190-062: 91 MHz
SCOPE levsine, 100 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
221 MHz
141 MHz
91 MHz