Specifications
Recorder 2
2-11
DC Voltage (VDC)
Ranges .................................................................... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading .................................................. 5000 counts
Accuracy .................................................................. ±(0.5 % +6 counts)
Normal Mode AC Rejection at
50 or 60 Hz ±1 % ........................................................ >60 dB
AC Voltage (VAC)
Ranges .................................................................... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading .................................................. 5000 counts
Accuracy
15 Hz to 60 Hz ..................................................... ±(1 % +10 counts)
60 Hz to 1 kHz ..................................................... ±(2.5 % +15 counts)
For higher frequencies the frequency roll-off of the Meter input starts affecting accuracy.
Normal Mode DC Rejection ..................................... >50 dB
AC+DC Voltage (True RMS)
Ranges .................................................................... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading .................................................. 5000 counts
Accuracy
DC to 60 Hz ......................................................... ±(1 % +10 counts)
60 Hz to 1 kHz ..................................................... ±(2.5 % +15 counts)
For higher frequencies the frequency roll-off of the Meter input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is larger than 5 % of full scale.
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges .................................................................... same as VDC, VAC, VAC+DC
Probe Sensitivity ...................................................... 100 μV/A, 1 mV/A, 10 mV/A, 100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy .................................................................. same as VDC, VAC, VAC+DC (add current probe or current shunt
accuracy)
Recorder
TrendPlot (Meter or Scope)
Chart recorder that plots a graph of min and max values of Meter or Scope measurements over time.
Measurement Speed ............................................... >5 measurements/s
Time/div ................................................................... 5 s/div to 30 min/div
Record Size (min, max, average) ............................ 19200 points
Recorded Time Span .............................................. 64 min to 546 hours
Time Reference ....................................................... time from start, time of day
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals