EasyManua.ls Logo

Fluke ScopeMeter 190 II Series - Using Average, Persistence and Glitch Capture

Fluke ScopeMeter 190 II Series
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
2-10
Meter Input (Banana Jacks)
Input Coupling ......................................................... DC
Frequency Response .............................................. DC to 10 kHz (-3 dB)
Input Impedance ..................................................... 1 MΩ (±1 %)//14 pF (±1.5 pF)
Max. Input Voltage ........................................... 1000 V CAT III
600 V CAT IV
(For detailed specifications, see “Safety”)
Meter Functions
Ranging ................................................................... Auto, Manual
Modes ..................................................................... Normal, Relative
General
DC Common Mode Rejection (CMRR) ................... >100 dB
AC Common Mode Rejection at 50, 60,
or 400 Hz ....................................................................... >60 dB
Ohms (
Ω
)
Ranges .................................................................... 500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ
Full Scale Reading
500 Ω to 5 MΩ .................................................... 5000 counts
30 MΩ ................................................................. 3000 counts
Accuracy ................................................................. ±(0.6 % +6 counts)
Measurement Current ............................................. 0.5 mA to 50 nA, ±20 % decreases with increasing ranges
Open Circuit Voltage ............................................... <4 V
Continuity (CONT)
Beep ........................................................................ <50 Ω (±30 Ω)
Measurement Current ............................................. 0.5 mA, ±20 %
Detection of shorts of .............................................. 1 ms
Diode
Maximum Voltage Reading ..................................... 2.8 V
Open Circuit Voltage ............................................... <4 V
Accuracy ................................................................. ±(2 % +5 counts)
Measurement Current ............................................. 0.5 mA, ±20 %
Temperature (TEMP)
With Optional Temperature Probe
Ranges (°C or °F) ................................................... -40.0 to +100.0 °
-100.0 to +250.0 °
-100.0 to +500.0 °
-100 to +1000 °
-100 to +2500 °
Probe Sensitivity ..................................................... 1 mV/°C and 1 mV/°F
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Related product manuals