Specifications
Introduction 2
2-3
Introduction
Specifications are subject to change without notification.
Oscilloscope
Isolated Inputs A, B, C and D (Vertical)
Number of Channels
Fluke 190-xx2 ...................................................... 2 (A, B)
Fluke 190-xx4 ...................................................... 4 (A, B, C, D)
Bandwidth, DC Coupled
Fluke 190-50x ...................................................... 500 MHz (-3 dB)
Fluke 190-2xx ...................................................... 200 MHz (-3 dB)
Fluke 190-1xx ...................................................... 100 MHz (-3 dB)
Fluke 190-062 ..................................................... 60 MHz (-3 dB)
Lower Frequency Limit, AC Coupled
with 10:1 probe .................................................... <2 Hz (-3 dB)
direct (1:1) ........................................................... <5 Hz (-3 dB)
Rise Time
Fluke 190-50x ...................................................... 0.7 ns
Fluke 190-2xx ...................................................... 1.7 ns
Fluke 190-1xx ...................................................... 3.5 ns
Fluke 190-062 ..................................................... 5.8 ns
Analog Bandwidth Limiters ...................................... 20 MHz, 20 kHz, and 10 kHz (varies according to version)
Input Coupling ......................................................... AC, DC
Polarity .................................................................... Normal, Inverted
Sensitivity Ranges
with 10:1 probe .................................................... 20 mV to 1000 V/div
direct (1:1) ........................................................... 2 mV to 100 V/div
Dynamic Range ....................................................... > ±8 div (<10 MHz)
> ±4 div (>10 MHz)
Waveform Positioning Range .................................. ±4 divisions
Input Impedance on BNC, DC Coupled
4-channel models ................................................ 1 MΩ (±1 %)//14 pF (±2.25 pF)
2-channel models ................................................ 1 MΩ (±1 %)//15 pF (±2.25 pF)
Vertical Accuracy ..................................................... ±(2.1 % + 0.04 range/div)
2 mV/div .............................................................. ±(2.9 % + 0.08 range/div)
Digitizer Resolution ................................................. 8 bits, separate digitizer for each input
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals