Performance Verification
Meter Tests 4
4-37
Table 4-18. Meter Volts AC Measurement Verification Points
Range 5502A output V ac Frequency Meter Reading
500.0 mV 500.0 mV
60 Hz 494.0 to 506.0
1 kHz 486.0 to 514.0
3 kHz >350.0
5.000 V 5.000 V
60 Hz 4.940 to 5.060
1 kHz 4.860 to 5.140
3 kHz >3.500
50.00 V 50.00 V
60 Hz 49.40 to 50.60
1 kHz 48.60 to 51.40
3 kHz >35.00
500.0 V 500.0 V
60 Hz 494.0 to 506.0
1 kHz 486.0 to 514.0
3 kHz >350.0
1100 V (1.1 kV) 1000 V
60 Hz 0.980 to 1.020
1 kHz 0.960 to 1.040
3 kHz > 0.700
Continuity Function Test
To test the continuity function:
1. Select the Test Tool setup.
2. Press
METER
.
3. Press
F1
to open the Measurement menu and select Continuity.
4. Connect the Test Tool to the 5502A as shown in Figure 4-13.
5. Set the 5502A to 20 Ω. Use the 5502A “COMP OFF” mode.
6. Listen to hear that the beeper is on.
7. Set the 5502A to 80 Ω.
8. Listen to hear that the beeper is off.
9. When you are finished, set the 5502A to Standby.
Diode Test Function Test
To do the diode test function:
1. Select the Test Tool setup.
2. Press
METER
.
3. Press
F1
to open the Measurement menu and select Diode.
4. Connect the Test Tool to the 5502A as shown in Figure 4-13.