EasyManuals Logo

Fluke ScopeMeter 190 II Series Service Manual

Fluke ScopeMeter 190 II Series
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #114 background imageLoading...
Page #114 background image
Calibration Adjustment
Final Calibration (Firmware: V10.9 and Lower) 5
5-29
4. Connect the Test Tool to the 5502A SCOPE output as shown in Figure 5-11.
Use the 50 Ω termination.
5. Set the 5502A to generate a sine wave 50 MHz / 0.5 V pp (mode LEVSINE)
at the SCOPE output.
6. Set the 5502A to operate (OPR).
7. Press
F3
to start the calibration.
8. Wait until the display shows calibration status
:READY.
9. Set the 5502A to standby (STBY).
10. Continue at the Input A LF-HF Gain section.
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
ADC
B
50 OHM e.g. PM9585 or
TRM50
USE 50 OHM
TERMINATION
AB
METER
50 OHM e.g. PM9585 or
TRM50
For 190-502 and -504 only
use TRM50!
PM9091
2 CHANN. + METER
4 CHANNELS
Perf-Ver-D2.eps
Figure 5-11. 5502A SCOPE Output to Test Tool Input A
Input A LF-HF Gain
To do the Input A LF-HF Gain calibration:
1. Connect the Test Tool to the 5502A as shown in Figure 5-11.
2. The display must show step CL 0654. If it does not, press
F2
or
F1
to select the first calibration step in Table 5-9.
3. Set the 5502A SCOPE output to source the signal required for the first
calibration point in Table 5-9.
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals