EasyManuals Logo

Fluke ScopeMeter 190 II Series Service Manual

Fluke ScopeMeter 190 II Series
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-14
Input B LF-HF Gain
To do the Input B LF-HF Gain calibration:
1. Press
F2
to select the first calibration step in Table 5-2.
2. Connect Ch. B of the Test Tool to the 5502A as shown in Figure 5-4.
3. Set the 5502A SCOPE output to source the signal required for the first
calibration point in Table 5-2 (CL 0674, Pos B Fast).
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-2.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Input C LF-HF Gain section.
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
ADC
B
USE 50 OHM
TERMINATION
AB
METER
PM9091
2 CHANN. + METER
4 CHANNELS
50 OHM e.g. PM9585 or
TRM50
For 190-502 and -504
use only TRM50!
50 OHM e.g. PM9585 or
TRM50
Perf-Ver-E2.eps
Figure 5-4. 5502A SCOPE Output to Test Tool Input B
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals