Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-12
Input A LF-HF Gain
To do the Input A LF-HF Gain calibration:
1. Connect Ch. A of the Test Tool to the 5502A as shown in Figure 5-3.
2. The display must show step CL 0654 (Pos A Fast). If it does not, then press
F1
or
F2
to select the first calibration step in Table 5-1.
3. Set the 5502A SCOPE output to source the signal required for the first
calibration point in Table 5-1.
4. Set the 5502A in operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-1.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Input B LF-HF Gain section.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals