Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
2-6
Automatic Scope Measurements
The accuracy of all readings is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific
accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe accuracy. At
least 1.5 waveform period must be visible on the screen.
General
Inputs ...................................................................... A, B, C and D
DC Common Mode Rejection (CMRR) ................... >100 dB
AC Common Mode Rejection at
50, 60, or 400 Hz .................................................... >60 dB
DC Voltage (VDC)
Maximum Voltage
with 10:1 probe ................................................... 1000 V
direct (1:1) ........................................................... 300 V
Maximum Resolution
with 10:1 probe ................................................... 1 mV
direct (1:1) ........................................................... 100 μV
Full Scale Reading .................................................. 999 counts
Accuracy at 4 s to 10 μs/div, Fluke 190-xx2
2 mV/div .............................................................. ±(1.5 % + 10 counts)
5 mV/div to 100 V/div .......................................... ±(1.5 % + 6 counts)
Accuracy at 4 s to 10 μs/div, Fluke 190-xx4
2 mV/div .............................................................. ±(3 % + 10 counts)
5 mV/div to 100 V/div .......................................... ±(3 % + 6 counts)
Normal Mode AC Rejection
at 50 or 60 Hz ........................................................ >60 dB
AC Voltage (VAC)
Maximum Voltage
with 10:1 probe ................................................... 1000 V
direct (1:1) ........................................................... 300 V
Maximum Resolution
with 10:1 probe ................................................... 1 mV
direct (1:1) ........................................................... 100 μV
Full Scale Reading .................................................. 999 counts
Accuracy, Fluke 190-xx2
DC coupled:
DC to 60 Hz .................................................... ±(1.5 % +10 counts)
AC coupled, low frequencies:
50 Hz direct (1:1) ............................................ ±(1.5 % + 10 counts) 0.6 %
60 Hz direct (1:1) ............................................ ±(1.5 % + 10 counts) 0.4 %
With the 10:1 probe the low frequency roll-off point will be lowered to 2 Hz, which improves the AC accuracy for low
frequencies. When possible use DC coupling for maximum accuracy.
AC or DC coupled, high frequencies:
60 Hz to 20 kHz .............................................. ±(2.5 % + 15 counts)
20 kHz to 1 MHz ............................................. ±(5 % + 20 counts)
1 MHz to 25 MHz ............................................ ±(10 % + 20 counts)
For higher frequencies the instrument’s frequency roll-off starts affecting accuracy.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals