Specifications
Automatic Scope Measurements 2
2-7
Accuracy, Fluke 190-xx4
DC coupled:
DC to 60 Hz ..................................................... ±(3 % +10 counts)
AC coupled, low frequencies:
50 Hz direct (1:1) ............................................. ±(3 % + 10 counts) 0.6 %
60 Hz direct (1:1) ............................................. ±(3 % + 10 counts) 0.4 %
With the 10:1 probe the low frequency roll-off point will be lowered to 2 Hz, which improves the AC accuracy for low
frequencies. When possible use DC coupling for maximum accuracy.
AC or DC coupled, high frequencies:
60 Hz to 20 kHz ............................................... ±(4 % + 15 counts)
20 kHz to 1 MHz .............................................. ±(6 % + 20 counts)
1 MHz to 25 MHz............................................. ±(10 % + 20 counts)
For higher frequencies the instrument’s frequency roll-off starts affecting accuracy.
Normal Mode DC Rejection ..................................... >50 dB
All accuracies are valid if:
The waveform amplitude is larger than one division
At least 1.5 waveform period is on the screen
AC+DC Voltage (True RMS)
Maximum Voltage
with 10:1 probe .................................................... 1000 V
direct (1:1) ........................................................... 300 V
Maximum Resolution
with 10:1 probe .................................................... 1 mV
direct (1:1) ........................................................... 100 μV
Full Scale Reading ................................................... 1100 counts
Accuracy, Fluke 190-xx2
DC to 60 Hz ......................................................... ±(1.5 % + 10 counts)
60 Hz to 20 kHz ................................................... ±(2.5 % + 15 counts)
20 kHz to 1 MHz .................................................. ±(5 % + 20 counts)
1 MHz to 25 MHz ................................................. ±(10 % + 20 counts)
For higher frequencies the instrument’s frequency roll-off starts affecting accuracy.
Accuracy, Fluke 190-xx4
DC to 60 Hz ......................................................... ±(3 % + 10 counts)
60 Hz to 20 kHz ................................................... ±(4 % + 15 counts)
20 kHz to 1 MHz .................................................. ±(6 % + 20 counts)
1 MHz to 25 MHz ................................................. ±(10 % + 20 counts)
For higher frequencies the instrument’s frequency roll-off starts affecting accuracy.
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges .................................................................... same as VDC, VAC, VAC+DC
Probe Sensitivity ...................................................... 100 μV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400 mV/A, 1 V/A, 10 V/A, and
100 V/A
Accuracy .................................................................. same as VDC, VAC, VAC+DC (add current probe or current shunt
accuracy)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals