EasyManua.ls Logo

Fluke ScopeMeter 190 II Series - Maintaining the Test Tool; About this Chapter; Cleaning the Test Tool; Storing the Test Tool

Fluke ScopeMeter 190 II Series
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-18
Input D LF-HF Gain
To do the Input D LF-HF Gain calibration:
1. Press
F2
to select the first calibration step in Table 5-4.
2. Connect Ch. D of the Test Tool to the 5502A as shown in Figure 5-6.
3. Set the 5502A SCOPE output to source the signal required for the first
calibration point in Table 5-4 (CL 0675, Pos D Fast).
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-4.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Input AB Position section.
ADC
B
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
4 CHANNELS
For 190-502 and -504
use only TRM50!
Perf-Ver-G2.eps
Figure 5-6. 5502A SCOPE Output to Test Tool Input D
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Related product manuals