EasyManua.ls Logo

Fluke ScopeMeter 190 II Series - Using Cursors on Spectrum Measurements

Fluke ScopeMeter 190 II Series
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
4-20
5. Set the 5502A to source a sine wave according to the first test point in
Table 4-8.
Because no 50 Ω termination is applied, the 5502A leveled sine wave output
amplitude will be twice the set value.
6. Observe the readings
A, B, C, and D and check that they are not outside the
range shown under the appropriate column.
7. Continue through the test points.
8. When you are finished, set the 5502A to 0 (zero) Volt and Standby.
Table 4-8. Phase Measurement Verification Points
Time base 5502A-SC... MODE Frequency Voltage
Input A, B, C, D
Reading … Deg
20 ms/div wavegen, sine, 1 MΩ 10 Hz 600 mVpp -2 to +2
200 ns/div levsine 1 MHz 300 mVpp -2 to +2
20 ns/div levsine 10 MHz 300 mVpp -3 to +3
Time Base Test
Proceed as follows to test the time base accuracy:
1. Connect the Test Tool to the 5502A as shown in Figure 4-6.
For the Fluke 190-502 and 190-504 you must use the Fluke TRM50 50 Ω
terminator.
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
ADC
B
50 OHM e.g. PM9585 or
TRM50
USE 50 OHM
TERMINATION
AB
METER
50 OHM e.g. PM9585 or
TRM50
For 190-502 and -504 only
use TRM50!
PM9091
2 CHANN. + METER
4 CHANNELS
perf-ver-d2.eps
Figure 4-6. 5502A Scope Output to Test Tool Input A
2. Set the 5502A to source an 8 ms time marker (MODE marker).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Related product manuals