Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-44
Probe Calibration
To meet full user specifications, you need to adjust the supplied red (R), blue (B),
gray (G) and green (V) VPS410-II 10:1 voltage probes for optimal response.
To adjust the VPS410-II probes:
1. Connect the red probe from the red Input A BNC to the banana jack. See
Figure 5-19.
Probe Cal
Probe Cal Ref.
st8416_t.eps
Figure 5-19. 10:1 Probe Calibration Connection
2. Press
A
and then
F3
to open the Probe on A menu.
3. Select Probe Type:
Voltage | Attenuation: 10:1.
4. Press
ENTER
.
5. Press
F3
- PROBE A.
6.
Press
F1
- PROBE CAL and follow the instructions shown on the display.