Calibration Adjustment
Calibration Procedure Steps 5
5-5
Equipment Required For Calibration
The primary source instrument used in the calibration procedures is the Fluke
5502A. If a 5502A is not available, you can substitute another calibrator that
meets the minimum test requirements.
Fluke 5502A Multi-Product Calibrator, including SC600 Oscilloscope
Calibration Option.
Stackable test leads (4x) as supplied with the 5502A (required for Test Tools
with banana jacks and 2 BNC oscilloscope inputs).
50 Ω Coax Cables (4x): use Fluke PM9091 (1.5 m, 3 pcs.) and PM9092
(0.5 m, 3 pcs.). For Test Tools with banana jacks and 2 BNC oscilloscope
inputs 2 Coax Cables are sufficient.
50 Ω feed through termination, Fluke TRM50 (4x for Test Tools with 4 BNC
oscilloscope inputs; 2x for Test Tools with banana jacks and 2 BNC
oscilloscope inputs). The use of model TRM50 is mandatory for
Fluke 190-502.
Male BNC to Dual Female BNC adapter (3x), Fluke PM9093/001.
Dual Banana Plug to Female BNC Adapter (1x), Fluke PM9081/001.
Calibration Procedure Steps
To do a complete calibration adjustment you must do all steps:
1. Select the Calibration Mode.
2. Do the Contrast Calibration Adjustment.
3. Do the Warming-Up and Pre-Calibration.
4. Do the Final Calibration for V11.10 and Later section or Final Calibration
(Firmware: V10.9 and Lower). If the installed firmware is V09.00, V10.00 or
V10.4x you must do the steps in the Final Calibration (Firmware: V10.9 And
Lower) section.
If the installed firmware is V11.10 you must do the steps in the Final
Calibration For V11.10 and later section.
5. Save the Calibration Data and Exit the calibration mode.
6. Do the probe calibration.
The following partial calibrations are allowed:
Contrast calibration, do the above-mentioned steps 1, 2, and 5. If during
normal operation the display cannot be made dark or light enough, or if the
display after a Test Tool reset is too light or too dark, you can do this
calibration.
Probe calibration, do the above-mentioned step 6. The probe calibration
matches the probe to the used input channel.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals