Calibration Adjustment
Final Calibration (Firmware: V10.9 and Lower) 5
5-37
Warning
To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
4. Set the 5502A to operate (OPR).
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-17.
8. Set the 5502A to Standby, and continue at the Input ABCD (AB) Position
section.
Table 5-17. Input ABCD Gain Calibration Points
Cal step UUT input value (5502A SCOPE, MODE volt, SCOPE Z 1 MΩ, 1 kHz)
CL 0660 300 mV
CL 0604 500 mV
CL 0637 none (5500 standby)
CL 0504 500 mV
CL 0624 none (5500 standby)
CL 0599 10 mV
CL 0600 25 mV
CL 0601 50 mV
CL 0602 100 mV
CL 0622 none (5500 standby)
CL 0662 2 V
CL 0605 1 V
CL 0606 2.5 V
CL 0607 5 V
CL 0664 20 V
CL 0608 10 V
CL 0609 25 V
CL 0610 50 V (set 5502A to OPR)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com