Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-42
NORMAL SCOPE
FLUKE 5502A CALIBRATOR
USE STACKABLE TEST LEADS
AS SUPPLIED WITH CALIBRATOR
AB
METER
2 CHANN. + METER
Perf-Ver-Z2.eps
Figure 5-18. Four-wire Ohms Calibration Connections
3. Set the 5502A to the first test point in Table 5-20. Use the 5502A “COMP 2
wire” mode for the calibration adjustments up to and including 100 kΩ. For
the higher values, the 5502A will turn off the “COMP 2 wire” mode.
4. Set the 5502A to operate (OPR).
5. Press
F3
to start the calibration.
6. Wait until the display shows the calibration status
:READY.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Save Calibration Data and Exit section.
Table 5-20. Ohm Gain Calibration Points
Cal Step UUT input Value (5500 NORMAL)
CL 0910 100 Ω
CL 0911 1 kΩ
CL 0912 10 kΩ
CL 0913 100 kΩ
CL 0914 1 MΩ
CL 0915 10 MΩ