Performance Verification
Scope Input A, B, C, D Tests 4
4-29
ADC
B
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
4 CHANNELS
For 190-502 and -504
use only TRM50!
perf-ver-g2.eps
Figure 4-10. 5502A Scope Output to Test Tool Input D
2. Reset the Test Tool to select the Test Tool setup.
3. Press
D
to turn Input D on.
4. Use
MOVE
to move the Input D trace zero to the center grid line.
5. Press
TRIGGER
and use
F1
to select Input D as trigger source.
6. Use
mV
RANGE
and
RANGE
to change the sensitivity range to select manual
sensitivity ranging and lock the Input D sensitivity range on 2 V/div.
7. Use
TIMEsns
to select the time base indicated in Table 4-15.
8. Set the 5502A to source the leveled sine wave given in the first row of
Table 4-14.
9. Adjust the 5502A output voltage until the displayed trace has the amplitude
indicated under the appropriate column of Table 4-15.
10. Verify that the signal is well triggered.
If not, press
TRIGGER
and use
F3
to enable the for manual
Trigger Level adjustment. Adjust the trigger level and verify that the signal is
triggered. The trigger icon (
) indicates the trigger level.
11. Continue through the test points.