Service Manual
5-8
1.
Press
F1
three times to select maximum darkness calibration Contrast
(CL 0100).
2. Press
F3
(CALIBRATE). The display shows a dark test pattern, see
Figure 5-2.
3. Use
to adjust the display to the maximum darkness where the
test pattern is only just visible.
4. Press
F3
to return to the softkey menu.
5. Press
F2
to select default contrast calibration Contrast (CL 0110):
6. Press
F3
(CALIBRATE). The test pattern shows on the display at default
contrast.
7. Use
to set the display to optimal (this setting becomes the
default) contrast.
8. Press
F3
to return to the softkey menu.
9. Press
F2
to select maximum brightness calibration Contrast (CL 0120):
10. Press
F3
(CALIBRATE). The display shows a bright test pattern.
11. Use
to adjust the display to the maximum brightness where the
test pattern is only just visible.
12. Press
F3
to return to the softkey menu.
13. Now you can either
Exit, if only the Contrast had to be adjusted. Continue at the Save
Calibration Data and Exit section.
or
Do the complete calibration. Press
F2
to select the next step
(Warming-Up) and continue.
hpp204.eps
Figure 5-2. Display Test Pattern
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Fluke ScopeMeter 190 II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 190 II Series and is the answer not in the manual?

Fluke ScopeMeter 190 II Series Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 190 II Series
CategoryTest Equipment
LanguageEnglish

Related product manuals