EasyManuals Logo
Home>GE>Monitor>Dash 3000

GE Dash 3000 Service Manual

GE Dash 3000
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #133 background imageLoading...
Page #133 background image
Troubleshooting: Get error logs
2000966-542D Dash 3000/4000/5000 6-17
Connect the PC to Dash patient monitor or MC network
NOTE
Connect the patient monitor and PC to a reliable power source
during the transfer process. Do not use battery power. A battery
failure during the transfer process may result in a transfer process
failure.
1. Connect one end of the category 5 crossover or network patch cable to
the PC network card.
2. Connect the other end of the category 5 crossover patch cable as
follows:
Connect to the ETHERNET port of the docking station if the
patient monitor is using a Dash Port 2 docking station.
Connect to the ETHERNET port of the patient monitor if no
docking station or network is used.
Connect to the MC network Ethernet tap if transferring logs
from other devises on the network. If you connect directly to the
MC network Ethernet tap, verify that the patient monitor or
docking station is also connected to the MC network Ethernet
tap.
3. Power on the PC.
4. Continue with the instructions, “Change the PC’s IP address” on
page 6-18.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals