Field replaceable units: Recommended checkout
2000966-542D Dash 3000/4000/5000 7-53
DAS assembly  “NBP 
calibration” on 
page 8-37.
 “ECG tests” on page 8-22.
 “Respiration tests” on page 8-25.
 “Temperature tests” on page 8-26.
 “Cardiac output tests (option)” on page 8-27.
 “Invasive blood pressure tests (option)” on 
page 8-27.
 “Pulse oximetry tests for GE Ohmeda SPO2 
oximeter” on page 8-30. Use this procedure if 
you installed a GE Ohmeda DAS assembly.
 “Pulse oximetry tests for Masimo SET SPO2” 
on page 8-32. Use this procedure if you 
installed a Masimo DAS assembly.
 “Pulse oximetry tests for Nellcor OxiMax SPO2” 
on page 8-33. Use this procedure if you 
installed a Nellcor OxiMax DAS assembly.
 “Noninvasive blood pressure tests” on page 8-
35.
 “Wireless antenna signal strength test (option)” 
on page 8-49. Use this procedure if the wireless 
option is installed.
 “Wireless LAN test (option)” on page 8-52. Use 
this procedure if the wireless option is installed.
 Refer to the CO2 chapter in the patient 
monitor’s operator manual to checkout the 
mainstream End-tidal CO2.
 To check out the sidestream End-tidal CO
2
, 
refer to the documentation that came with the 
module.
 “Electrical safety tests” on page 8-4.
Recommended calibration, electrical safety tests, and checkout procedures
Replacement 
procedure
Calibration 
tests
Checkout procedures Electrical safety tests