Field replaceable units: Recommended checkout
7-54 Dash 3000/4000/5000 2000966-542D
NBP assembly  “NBP 
calibration” on 
page 8-37.
 “Noninvasive blood pressure tests” on page 8-
35.
 “Electrical safety tests” on page 8-4.
Wireless LAN  None  “Patient monitor power-up tests” on page 8-21.
 “Network test (option)” on page 8-46.
 “Wireless antenna signal strength test (option)” 
on page 8-49.
 “Wireless LAN test (option)” on page 8-52.
Wireless card None  “Wireless antenna signal strength test (option)” 
on page 8-49.
 “Wireless LAN test (option)” on page 8-52.
Power supply 
assembly
None  “Patient monitor power-up tests” on page 8-21.
 “Wireless antenna signal strength test (option)” 
on page 8-49. Use this procedure if the wireless 
option is installed.
 “Wireless LAN test (option)” on page 8-52. Use 
this procedure if the wireless option is installed.
CPU  “End-tidal 
CO2 test 
(option)” on 
page 8-44.
“NBP 
calibration” on 
page 8-37.
 “Patient monitor power-up tests” on page 8-21.
 “Analog output and defibrillator synchronization 
tests” on page 8-40.
 “Graph or print tests (option)” on page 8-44.
 “Display test” on page 8-45.
 “Speaker test” on page 8-46.
 “Network test (option)” on page 8-46.
 “Wireless antenna signal strength test (option)” 
on page 8-49. Use this procedure if the wireless 
option is installed.
 “Wireless LAN test (option)” on page 8-52. Use 
this procedure if the wireless option is installed.
 “TRAM-rac 2A module housing peripheral 
device test (option)” on page 8-54.
Writer, writer flex None  “Patient monitor power-up tests” on page 8-21.
 “Graph or print tests (option)” on page 8-44.
 “Wireless antenna signal strength test (option)” 
on page 8-49. Use this procedure if the wireless 
option is installed.
 “Wireless LAN test (option)” on page 8-52. Use 
this procedure if the wireless option is installed.
Recommended calibration, electrical safety tests, and checkout procedures
Replacement 
procedure
Calibration 
tests