Preventive maintenance: Cleaning and disinfecting the patient monitor
2000966-542D Dash 3000/4000/5000 5-5
Impact or results of improper cleaning products and processes
The following consequences can occur if you do not follow the cautionary
guidelines when cleaning the equipment:
Product discoloration.
Melting, dulling, or distorting of the device case.
Brittle and breaking device case.
Blurring the lettering on label text.
Equipment malfunction or failure.
Void warranty.
Cleaning products to avoid
Cleaning products known to cause the types of problems listed previously
include, but are not limited to:
Sani-Cloth
®
Wipes
Ascepti
®
Wipes
HB Quat
®
Clorox
®
Wipes (they do not contain bleach)
Over-the-counter detergents (e.g., Fantastic
®
, Tilex
®
, etc.)
Products that contain active ingredients and solutions similar to these
products should be avoided.
Storage
Always remove batteries when the device is not in use (even for short
periods of time).
Store in a dry well-ventilated area.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable