Functional and electrical safety checks: Functional Checkout procedures
2000966-542D Dash 3000/4000/5000 8-27
Cardiac output tests (option)
Equipment
A multiparameter patient simulator
Cardiac output cable adapter
Cardiac output simulator
Connections
1. Connect the cardiac output cable adaptor to the TEMP/CO connector
of the patient monitor.
2. Connect the cardiac output adapter to the cardiac output simulator.
3. Verify a cardiac output parameter window appears on the patient
monitor.
Cardiac output test procedures
1. On the patient monitor, select CO Parameter Box.
2. Set up the patient monitor:
Set AUTO MODE: to ON.
Set INJECT TEMP: to BATH.
Set SIZE: to 7.
Set INJECT VOL: to 10CC.
Set COMPUTATIONAL CONSTANT: to 0.540.
3. Set cardiac output cable adapter or simulator to .
4. Turn the simulator on. Select cardiac output menu corresponding to
the injectate temperature set in step 3 at 5 liters/minute
5. When the patient monitor screen displays the message, INJECT
WHEN READY, press appropriate button on simulator to inject.
6. When computing is complete, the CO reading should be
approximately 37°C at 5 liters of blood per minute ± 5%.
Invasive blood pressure tests (option)
Equipment
A multiparameter patient simulator
Dual BP cable (2005772-001 or equivalent)
The invasive blood pressure (BP) tests provide a method of verification
for all BP connectors (BP1, BP2, BP1/3 and BP2/4) of a patient monitor
equipped with this optional function.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals