Troubleshooting: Change internet address
6-12 Dash 3000/4000/5000 2000966-542D
Change internet address
The internet address is an identification number assigned to each device
on the optional CARESCAPE Network. If the internet address has been
corrupted, changing it must be done in Boot Code using a unique
password. Contact your sales/service representative and provide them
with the serial number and internet address of the unit to obtain a
password.
WARNING
Lost Data. Duplication of an internet address on a
network will cause lost data. If you change the factory-
assigned internet address, you must first record all other
internet addresses used on your network to avoid
duplication.
Activate the Boot Loader program as follows:
1. Hold down NBP Go/Stop (or NBP Auto on Dash 5000) and ZERO
ALL (FUNCTION) on the front panel.
2. Press and release the Trim Knob control.
3. Keep holding NBP Go/Stop (or NBP Auto) and ZERO ALL
(FUNCTION) until the Boot Loader information appears on the
display.
4. Select Service Menu > Option Menu > Change Internet Address
to acquire information required for password.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals